Skip to content

Maxwell® CSC Instrument

by Promega
Login to view price.
Size
  • Overview
  • Protocols
  • Specifications
  • Resources
  • Manufactured under cGMP

    The Maxwell® Clinical CSC automated nucleic acid extraction system is an in vitro diagnostic medical device specifically designed for clinical laboratories processing multiple samples for critical downstream assays. Manufactured under cGMP, the consistent performance of the Maxwell® CSC Instrument assures extraction of high-quality DNA or RNA from every run.

    Currently only available in the United States, Canada, China, Korea, Austria, Belgium, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, Poland and the United Kingdom.

    .

    12589CBLargeCSC11
    Records displayed on tablet

    Maintain Electronic Records Easily

    Maxwell® CSC makes maintaining electronic records easy with complete tracking of every run and reports that are portable and LIMS-compatible. The instrument meets the 21 CFR Part 11 requirements for electronic records (user authentication and authorization, data integrity and protection, electronic signatures and audit trails) when used with the appropriate laboratory workflow.

     

    Fewer cross-contamination worries with a paramagnetic particle mover

    The Maxwell® CSC is a magnetic particle mover, not a liquid handler. It delivers minimal risk of cross-contamination because no liquid handling or splashing happens during processing of the samples.

    12624MC-W-a
    12624MC-W-b
    12624MC-W-c

    The Maxwell® CSC makes it easy to set up and run samples

    Scanning Maxwell CSC Blood DNA Kit

    Track Samples Easily


    The integrated barcode reader records important information such as sample ID, kit ID and kit lot number.

    Running a protocol from tablet

    Easy to Use


    The intuitive touch-screen interface and preprogrammed protocols lets you easily set up your purification run.

    Adding Samples to Cartridges

    Reduce Processing Time


    The prefilled reagent cartridges are ready to use. Simply add your sample, place the cartridges in the Maxwell® CSC and press "Start".

    Instrument Specifications

    Processing Time: 40–60 minutes (depending on sample type and method)

    Number of Samples: Up to 16

    Weight: 24.2lb (11kg)

    Dimensions: 13W × 13.6D × 11.8H inches (330.2 × 345.2 × 299.7mm)

    Power Requirements: 95–240VAC, 50–60Hz, 1.0A

    Fuse: 250VAC, 2.6A, time-lag fuse (AC250V, T2.5AL, 5 × 20mm)

    UV Bulb: Average lifetime approximately 6,000 hours, length 135.9mm, diameter 16mm, 4W, 0.17A current, 29V, Spectral Peak F 253.7, UV output 0.9W

    Instrument Standard Components

    Maxwell CSC Component A
    maxwell csc components B through I

    A. Maxwell® CSC Instrument

    B. Tablet PC preloaded with Maxwell® CSC User Interface

    C. USB cable for connection of the Maxwell® CSC Instrument to the Tablet PC

    D. Power Cable for Maxwell® CSC Instrument

    E. Power Cords for Tablet PC

    F. Maxwell® CSC Deck Tray

    G. Bar Code Reader

    H. Tablet PC Holder

    I. USB cable - Maxwell® CSC Instrument to the Tablet PC

    Software CD for the Maxwell® CSC Instrument

  • Protocols

  • Specifications

    Certificate of Analysis

    See all certificates for AS6000

    Use Restrictions

    For In Vitro Diagnostic Use. This product is only available in certain countries.

    Storage Conditions

    DD

    Patents and Disclaimers

    U.S. Pat. Nos. 7,721,947 and 7,891,549 and other patents and patents pending.